pH meter, pH controller Model MDA-PH Measure range pH: 0-14 pH, ±0.02pH ORP: -1000 ~1000mV, ±1mV Measuring medium Liquid Input Resistance ≥1012Ω Temp compensation Manual/ Auto te.
Post Views: 182
The pH measurement system operates through a sophisticated yet reliable electrochemical process. It consists of a rugged pH electrode (sensor) and an intelligent transmitter (controller/display).
The pH electrode generates a millivolt-level electrical signal that is directly proportional to the hydrogen ion activity in the solution. This signal is exceptionally weak and susceptible to interference. The transmitter performs two critical functions: first, it amplifies this weak analog signal; second, it converts it into a stable digital value. This value is then calculated against the internal calibration curve (typically set at pH 4.0, 7.0, and 10.0 buffer solutions) to provide an accurate, temperature-compensated pH reading on the clear display. The two-wire or four-wire loop-powered design also allows for seamless integration into modern process control systems.
| Model | MDA-PH |
| Measure range | pH: 0-14 pH, ±0.02pH |
| Measuring medium | Liquid |
| Input Resistance | ≥1012Ω |
| Temp compensation | Manual/ Auto temperature compensation |
| Temperature Range | -10~130℃, NTC10K or PT1000 |
| Communication | RS485, Modbus-RTU |
| Signal output | 4-20mA, maximum loop 750Ω, 0.2%FS |
| Power supply | 220V±10%,24V±20%,50Hz/60Hz |
| Relay output | 250V, 3A |